GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-7
Be sure all USB devices (printers only) are connected and turned on (except for USB Flash
d
evices). If a printer is plugged in at this time, it is automatically installed.
Save all Service Presets if you are connected to InSite to restore remote connectivity af
ter
something is
completed.
Save all User Presets.
NOTE: After a full or partial system software load, be sure to check the system date, time and time zone are
set correctly. Confirm the Windows automatic Daylight Saving Time feature is turned off (Utility ->
System -> General -> Date/Time -> Time Zone tab -> Automatically Adjust Clock checkbox).
System and Application software disables the Run window normally accessed from Start -> Run, and
the mouse right-click (left set key) feature.
8-3-5-2 Preparation for Software Loading
8-3-5-2-1 Recording important settings and parameters (optional)
If the system has not been deployed into a customer site yet (i.e. the system has been in the factory
default and there has not been patient data), this step can be skipped.
8-3-5-2-1-1 Overview
See the following pages to record data.
These subsections include descriptions for recording/setting up data for the following screens.
With the system Powered On, follow the procedures below to document basic system information and
back up Presets, and if required, Patient Archive and Patient Images prior to software load.
Utility -> Connectivity
- TCP/IP (see: “Connectivity - Recording the TCP/IP settings”).
- Device (see: “Connectivity - Recording the AE Title and Port settings”).
- Miscellaneous (see: “Connectivity - Recording Miscellaneous settings”).
Utility -> Admin
- System Admin (see: “Admin - Recording the Software Option Keys”).
These subsections include descriptions for recording/setting up data for the following screens.
Hardcopy forms are provided here for convenience, but content will be included if selected during
Backup process. Backing up this information is only required if an upgrade specifies the information be
backed up or as an extra precaution against loss due to a corrupt back-up disk.
Utility -> Connectivity
- Device (see: “Connectivity - Recording Device Destination settings”).
- Service (see: “Connectivity - Recording Service settings”).
- Dataflow (see: “Connectivity - Recording Dataflow settings”).
- Buttons (see: “Connectivity - Recording the Print Key Assignments”).
- Removable Media (see: “Connectivity - Service settings”).
Utility -> Admin
- Users (see: “Admin - Users”).
Utility -> System
- Backup/Restore (see: “System - Data Store Management”).
- Peripherals (see: “System - Recording Peripheral settings”).
- About (see: “System - Recording software and hardware versions”).