GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-11
7-4-3-4 Hardware configuration error
7-4-3-4-1 Cause
Hardware(s) not detected as valid system. System will boot up but will not allow scanning.
7-4-3-4-2 What to do
1.Press OK, and shutdown the system and reboot.
2.If same issue persists, then shutdown the system, system breaker.
7-4-3-4-3 Log message of the Invalid hardware error
Example
Error ; acqHWCAP(1776); ********************************************
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCap(1776); *** HW RACK CONFIGURATION INVALID!!! ***
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCAP(1776); ********************************************
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCAP(1776); *** Board FuncRevs Provided ***
Error ; acqHWCAP(1776); *** ----------------------- ***
Error ; acqHWCAP(1776); *** BF_FuncRev1D (qty=1)
Error ; acqHWCAP(1776); *** COMEXP_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** CPS_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** GFS_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** PIL_FuncRev1A (qty=1)
Error ; acqHWCAP(1776); *** ***
Error ; acqHWCAP(1776); ********************************************
Info ; DFEManager(1776); HandleRunFlag: setting RunDfe and RunBlock flags to true!
Error ; EAErrorHandler(1776); Error from acqHWCAP: Invalid Hardware Rack Configuration - Check
logs. Severity : FatalHW