GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-99
4.) Note : If following windows pop up without changing any parameters (i.e. simply browsing through
the VPD contents), then press “NO” before continuing.
Figure 8-110 Question windows
8-21-6-2 Initializing EEPROM
When a new EEPROM is attached to a system, it will require initialization.
1.) Select “COMEXPRESS” from Device Grid (right pane) .
Figure 8-111 COMEXPRESS
2.) Select “Load Template” from right pane.
3.) Select “ComexpMaster.VPD”.
File is found under c:\Scanner\target\resources\acquisition\Housekeeping
.
Figure 8-112 Load Tmplate