GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 4 - Functional Checks 4-26
4-3-7-6 M-Mode
1.) Press B Mode Button.
2.) Press M Mode Button.
3.) Verify that system scan B/M with the activated probe.
Figure 4-26 M-Mode
4-3-7-7 B-Flow Mode
NOTE: Make sure to select probe which supports B-flow
1.) From the touch panel, select Utility -> Imaging -> General.
2.) Change BF/BFC Button to “BF”.
Figure 4-27 BF Setting
3.) Press B-Flow Button.