8-124 Section 8-29 - Replacement of the Peripherals
8-29-1-2-4 Installation Procedure
1.) Parts to be installed in reverse order on removal.
8-29-1-2-5 Functional Check
Test 1: Print an image by pressing the button P1 on the operator panel. (P1 key should be
configured correctly)
Verify1: - Verify that image can be printed.
Verify2: - Verify that printed image has no distortion as compared the image on screen.
8-29-2 Color Printer
8-29-2-1 Color printer on Mid Console
8-29-2-1-1 Manpower
1 person, 1 hour
8-29-2-1-2 Tools
Philips screwdriver
8-29-2-1-3 Removal Procedure
1.) Pull out the cabinet.
Figure 8-141 Remove the cabinet