EasyManua.ls Logo

GE LOGIQ S7 Pro - Replacement Around ACQ Box

GE LOGIQ S7 Pro
435 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-70 Section 8-16 - Replacement around ACQ Box
Section 8-16
Replacement around ACQ Box
NOTE: This section covers replacement of components related to ACQ Box - PIL assy, GBF128/64 Assy, GFS
Assy, SOM w/ 4GB Memory, HDD, CPS Assy, CBP Assy, CSI Assy, FAN w/ Cable Assy, DC4D Assy
and DVR Assy (for R1.x.x only)
8-16-1 Manpower
One person, 15 minutes for each parts.
8-16-2 Tools
Phillips screwdriver.
8-16-3 ACQ Box Access
1.) Remove Rear Cover.
2.) Remove Power Cord Fix Bracket.
Figure 8-67 Power Card
3.) Remove Cable Cover Bracket.
Figure 8-68 Cable Cover Bracket