EasyManua.ls Logo

GE LOGIQ S7 Pro - Removal Procedure - Separating CWD Board from PIL

GE LOGIQ S7 Pro
435 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-75
8-16-4-2 Removal Procedure - Separating CWD Board from PIL
Philips screw-driver required to add this part.
1.) Locate CW connector on PIL Board.
Figure 8-74 PIL Board
2.) Remove three screws from back of CW connector board.
Figure 8-75 CW connector board
8-16-4-3 Handling EEPROM on GFS Board
EEPROM on GFS board contains vital informations.
A.) System Serial Number that is matched against stored software option key.
B.) ComExpress Functional REV that is monitored for sub-system compatibility.
CWD