8-90 Section 8-20 - Replacement of Caster
2.) Remove Nut.
Figure 8-95 Removable Nut of Caster
8-20-4 Installation Procedure
Parts to be installed in reverse order of removal.
8-20-5 Functional Check
!! CAUTION:
Proceed with extreme caution and avoid system tipping over. The system weighs approximately
90kg.
Test Refer to Note
Mechanical Parts Functional Check Section 4-3-2 Only Swivel/Brake Lock Caster applies