GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 4 - Functional Checks 4-25
5.) Revert to original setting in Utility -> Imaging -> General “Simultaneous”.
Figure 4-24 Pulse Doppler Mode
4-3-7-5 B/CF/PW Mode
1.) Press B Mode button.
2.) Press PW Mode Button.
3.) Press CF Mode Button.
4.) Verify that system scan with B/CFM/PWD Modes.
Figure 4-25 B/CF/PW Mode