GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-9
7-4-3 System Warning/Errors and Logs
7-4-3-1 Temperature warning
7-4-3-1-1 Cause
CPU Temperature exceeded threshold (95°C).
7-4-3-1-2 What to do
1.Shut down the system and Clean the filter.
7-4-3-2 Temperature exceeded threshold
7-4-3-2-1 Cause
CPU Temperature exceeded threshold (100°C) and therefore, system must be shut down immediately.
7-4-3-2-2 What to do
1.Shut down the system and Clean the filter.
7-4-3-2-3 Log message of the high temperature
Example
Error ; acqHouseKeeping(****); CPU Temperature Sensor BEP_CPU_TEMPERATURE = 97.00
degrees C is outside tolerance limits. UpperLimit = 100.00 LowerLimit = -10.00