GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-11
10-5-7 Probe Maintenance
10-5-7-1 Probe Related Checks
10-5-7-2 Basic Probe Care
The Basic User Manual and/or care card provides a complete description of probe care, maintenance,
cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes.
Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of
LOGIQ™ S7 Expert/Pro for more details. Failure to follow these precautions can result in serious injury
and equipment damage. Failure to properly handle or maintain a probe may also void its warranty.
Any evidence of wear indicates the probe cannot be used.
Do a visual check of the probe pins and system sockets before plugging in a probe.
10-5-7-3 Basic Probe Cleaning and/or Disinfection
Refer to the Basic User Manual of LOGIQ™ S7 Expert/Pro for details on cleaning.
Table 10-10 Probe Related Checks
Step Item Description
1 Probes
Thoroughly check the system probe connectors and remove dust from inside the connector
sockets if necessary. Visually check for bent, damaged or missing pins.
2 Probe Holder
Clean probe holders (they may need to be soaked to remove excess gel).
CAUTION
!! CAUTION:
Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s
warranty. DO NOT soak or wipe the lens with any product not listed in the LOGIQ™ S7 Expert/
Pro Basic User Manual and/or care card. Doing so could result in irreparable damage to the
probe and/or LOGIQ™ S7 Expert/Pro system.
CAUTION
!! CAUTION:
Follow the Care Card instructions supplied with each probe (inside the transducer boxes)
for disinfectants and gels that are compatible with the surface material of the probes.
CAUTION
!! CAUTION:
To help protect yourself from blood borne diseases, when cleaning and handling probes,
wear approved, non-allergenic disposable gloves.
NOTICE
!! NOTICE:
Disinfect a defective probe before you return it. Be sure to tag the probe as being disinfected.
OK OKOK
CAUTION
CAUTION:
Please be aware of the
sensitive probe head.
TAKE EXTREME CARE!
NEVER place or store a
probe on its scan head!
When
disinfecting a probe, ensure
that there is sufficient space
between the probe and the

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S7 Pro and is the answer not in the manual?

GE LOGIQ S7 Pro Specifications

General IconGeneral
BrandGE
ModelLOGIQ S7 Pro
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals