GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-101
8-21-6-3 Modifying contents of EEPROM (ComExpress VPD)
System reads off Console Serial Number from EEPROM and match that against installed option string.
Therefore, without having correct serial number in EEPROM (ComExpress VPD).
1.) Select “COMEXPRESS” from Device Grid (right pane).
Figure 8-114 COMEXPRESS
2.) Double Click value for “System Serial Number”.
3.) Type in console’s serial number.
4.) Click “Write Dirty Fields” button on right pane.
NOTE: This procedure can be applied to other components with VPD, for example, for modifying GFS FuncRev
value. However, it is unlikely need for such event arises, as these VPD components are hardware-
integrated into boards.
8-21-7 Functional Check
Test Refer to Note
Board Diagnostics Section Section 4-4