GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
10-22 Section 10-8 - Electrical Safety Tests
10-8-2 GEHC Leakage Current Limits
The following limits are summarized for UL60601-1 (For USA), IEC 60601-1 Medical Equipment Safety
Standards and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical
electrical equipment.
Measurement limits per UL60601-1 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
Table 10-11 Enclosure Leakage Current Limits - Accessible Surfaces not protectively earthed
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.1mA* 0.1mA***
Close Protective earth and close neutral with reverse polarity 0.1mA* 0.1mA***
Open Protective earth and close neutral with normal polarity 0.3mA** 0.5mA***
Open Protective earth and close neutral with reverse polarity 0.3mA** 0.5mA***
Open neutral and close Protective earth with normal polarity 0.5mA* 0.5mA***
Open neutral and close Protective earth with reverse polarity 0.5mA* 0.5mA***
Table 10-12 Earth Leakage Current Limits
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.3mA** 0.5mA***
Close Protective earth and close neutral with reverse polarity 0.3mA** 0.5mA***
Open neutral and close Protective earth with normal polarity 1mA* 1mA***
Open neutral and close Protective earth with reverse polarity 1mA* 1mA***
Table 10-13 Type BF Patient Leakage Limits - Non-Conductive Surface
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.1mA* 0.1mA***
Close Protective earth and close neutral with reverse polarity 0.1mA* 0.1mA***
Open Protective earth and close neutral with normal polarity 0.5mA* 0.5mA***
Open Protective earth and close neutral with reverse polarity 0.5mA* 0.5mA***
Open neutral and close Protective earth with normal polarity 0.5mA* 0.5mA*
Open neutral and close Protective earth with reverse polarity 0.5mA* 0.5mA***