GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 4 - Functional Checks 4-8
4-3-2-3 OPIO Swivel Lock
Figure 4-7 OPIO Swivel
1.) Push the swivel button and hold it. (see: Figure 4-7)
2.) Rotate OPIO (while holding button).
3.) Verify OPIO is free to rotate.
NOTE: OPIO can rotate approximately 30 degrees to each side.
4.) Let go of Swivel Lock button (disengage).
5.) Verify OPIO does not rotate.
NOTICE
!! NOTICE:
Do NOT attempt to apply excessive rotating force to OPIO. OPIO Swivel brake is designed to hold
OPIO in desired position during normal use, and not meant to completely fix the OPIO position.