Chapter 7 - Diagnostics/Troubleshooting 7-13
7-4-3-5-3 Log message of System error
Log message format
“Error ; EAErrorHandler(****); Error from” + ErrorModuleName + ErrorMessage
Log message of System error
Example1
Exception Type
GFI Hardware Exception
GFI Setup Exception
GFI Buffer Exception
GFI Processing Exception
Sub Type Code
Code Error name description
31 eSwHardwareException_ErrorHandler_GFIHardwareError Hard (BF, GFI, CPS) error
167 eSwProcessingException_TsLevelOk_Timeout no output for CPS HV (High voltage)
Error description
Message description
GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_HVLV). Severity :
FatalHW
CPS error
GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_POWER). Severity :
FatalHW
CPS error
GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_STOP). Severity :
FatalHW
CPS error
GFE Ch0 empty-GFE Ch1 empty-BF_ERROR_A . Severity : FatalHW BF board error
GFE Ch0 empty-GFE Ch1 empty-BF_ERROR_B . Severity : FatalHW BF board error
GFE Ch0 empty-GFE Ch1 empty-BF_ERROR_C . Severity : FatalHW BF board error
GFE Ch0 empty-GFE Ch1 empty-EMERGENCY_ERROR . Severity : FatalHW GFS board error
scanctrl ScanControl::startScan: Tx Power Supply is not ready. HV setup
timeout after 4127 ms, ScanSequencer NOT started!. Severity : NonFatalSW
no output for CPS HV (High voltage)