GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-72 Section 8-16 - Replacement around ACQ Box
8-16-4 Removal Procedure - PIL, GBF128/64, GFS, CPS
NOTE: Before working with ACQ box, be sure to remove probes.
1.) Get ACQ Box access.
2.) Remove 9 screws.
Figure 8-70 ACQ Box
3.) Remove EMI Bracket to expose PIL, GBF128/64, GFS, and CPS
4.) Draw PIL, GBF128/64, GFS, CPS out of the NEST Frame
Figure 8-71 Separating Nest Assy