GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-60 Section 8-13 - Replacement of OPIO and Related Parts
8-13-3-8 Removal Procedure - OP Panel Main PWA Assy
1.) Remove knobs from OPIO.
2.) Disconnect 7 cables marked below.
3.) Remove 27 screws around PWA board.
4.) Slowly lift board from OP Panel.
Figure 8-53 Main PWA Assy
8-13-3-9 Removal Procedure - Joystick Assy
1.) Disconnect 2 cables marked below.
2.) Remove 3 screws from Joystick Assy.
3.) Slowly lift board form OP Panel.
Figure 8-54 Joystick Assy