GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
5-26 Section 5-8 - Exiting to Windows Desktop
Section 5-8
Exiting to Windows Desktop
LOGIQ™ S7 Expert/Pro backend software operates on embedded version of Windows XP
®
operating
system. To access some of the functions, service personnel occasionally require to enter Windows
desktop to access files, folders or applications.
5-8-1 Exiting to Windows Desktop
Tools Required: Service Dongle
1.) Wait for system to boot.
2.) Connect Service Dongle to any of the USB ports available.
3.) Tap power button once.
4.) If Service Dongle is correctly recognized by the system, U/I popup window provides selection
“EXIT”.
Figure 5-10 SYSTEM - EXIT
5.) On “Maintenance Access” window, type in service password (user name not required).
NOTE: Service Password is rotating password that changes on 6 months period. Only the GE service
personnel or GE authorized dealers have access to maintenance password.
NOTE: Maintenance Password is function of date/time. Therefore, if BIOS settings differ from actual date/time,
then the maintenance password may not be recognized.