GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 4 - Functional Checks 4-21
Figure 4-20
6.) Connect any probe. (if not connected)
7.) Go to Live CF Mode. (Press “CF” Mode button, unfreeze.)
8.) Verify both B button and CF button are lit GREEN.
9.) Turn “GAIN” encoder around CF Mode button and verify ROI gain change.
10.)Operate TGC sliders and verify B image gain change.
11.)Press “COMMENT” key.
12.)Type in arbitary strings from A/N Keyboard. Verify keyboard is functional.
4-3-5-2 DVD Drive Test
1.) Insert Blank CD-R or DVD-R Media into Optical Drive.
2.) Select UTILITY -> SYSTEM -> Backup Restore.
3.) Select “CD/DVD” in “MEDIA” list.
4.) Select “User Define Configuration checkbox” in Backup and Press “Backup”.
5.) Start backup into Media by pressing “OK” Button.
6.) Verify “Finished OK” message appears.
4-3-5-3 LAN Port Test
1.) Plug the Ethernet Cable to connect local area network.
2.) Verify that Local area network connection is recognized.