GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-61
8-13-3-10 Removal Procedure - Joystick and Encoder on PWA
1.) Remove OP Panel Main PWA Assy.
2.) Remove “Rubber Sheet” on reverse side of PWA Assy.
3.) Loosen nut.
4.) Disconnect Connector (on reverse side).
5.) Remove part.
Figure 8-55 Joystick and Encoder