Loading...
GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-29
8-3-8 Software Installation from Common Service Desktop
NOTE: This procedure requires PC with LAN connection capable of reaching to to GE Network
8-3-8-1 Involking Software Reload
1.) Select [SERVICE] from [UTILITY] Menu
2.) Select “GE Service” and enter Password
3.) Verify entering to Common Service Desktop (CSD)
4.) Select “Disruptive Mode” under Common Utility, in [Utilities] Tab
5.) Acknowledge “Disruptive Mode” on Console Screen