GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-63
Section 8-14
Replacement of the Caps for Hardkeys
8-14-1 Manpower
One person, approx. 1 minute/cap.
8-14-2 Tools
Small-sized slotted screwdriver or tweezers.
8-14-3 Removal Procedure
1.) By means of a small slotted screwdriver, carefully push against the hardkey cap.
2.) Lift the cap, until it is completely loosened from its base.
3.) Place the new hardkey cap down until it snaps into position.
NOTE: Each of the key have “Slit” where flathead screwdriver can be inserted for easy removal.
Figure 8-58 Slit for removal
8-14-4 Installation Procedure
1.) Parts to be installed in reverse order of removal.
Figure 8-57 Push against the circle cap and lift it
“Slit” for removal