EasyManua.ls Logo

GE LOGIQ S7 Pro - System Integration Checks; OPIO Test

GE LOGIQ S7 Pro
435 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 4 - Functional Checks 4-20
4-3-5 System Integration Checks
4-3-5-1 OPIO Test
Figure 4-18 OPIO
1.) Power ON the system.
2.) Check backlight on AN Keyboard, B Buttons, Freeze Key.
3.) Select UTILITY in touch panel.
4.) Turn encorder knobs as shown in Fig 4-18.
5.) Verify backlight luminance (intensity) changes when knobs are turned.
Figure 4-19 OPIO
Utility

Table of Contents

Related product manuals