GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-33
Section 8-4
Software and Functional Checks after Installation/Upgrade Procedure
8-4-1 Functional Check
Section 8-5
Settings - Backup and Restore
This section briefly describes Image Setting backup and restore function.
For detailed documentations and warnings, refer to Chapter.16, Section “System/Backup and Restore
Preset Menu” in the Basic User Manual.
8-5-1 General Information
The Backup/Restore function enables the user to:
Copy/Restore the patient archive.
Copy/Restore the system configuration. The Copy/Restore system configuration feature enables
the user to configure several units with identical presets, providing that the units have the same
software version.
8-5-1-1 Supported Media
Depending on the system, you can use either a CD-R, DVD-R, USB Flash Drive, or USB Hard Disk for
system backup/restore. System does NOT support CD-RW/DVD+R/DVD+RW.
For the sake of simplicity, we have used the CD-R in the examples in this section.
8-5-1-2 General Notes
NOTE: When you restore backup data from the Utility menu, the LOGIQ S7 Expert/Pro application usually
restarts automatically when the restoring is complete.
NOTE: To perform backup and restore procedures, you must login with administrator privileges.
8-5-2 Backup Procedures
Back up patient data AFTER you’ve archived (via EZBackup/EZMove) images so that the pointers to
the patients images reflect that the images have been moved to removable media and are no longer on
the hard drive.
1.) Insert a media into the drive or USB device into a USB port.
NOTE: About formatting media, See ‘Formatting removable media’ on Basic User Manual for more information.
2.) In the patient screen, select the dataflow Local Archive - Int. HD.
3.) On the touch panel, press Utility.
Table 8-18
Item Test Procedure Reference Note
Power ON 4-3-3 Power On/Off
System Info 4-3-4 System Information
Mode Transition 4-3-7 Mode Transition Checks

Table of Contents