GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-65
Section 8-15
Replacement of Key Caps (by special native language keys)
8-15-1 Manpower
One person, 30 minutes.
8-15-2 Tools
• Regular Phillips screwdriver
• A/N key assembly removal tool (shipped with AN keyboard)
• Screwdriver (Diameter : 3mm)
Figure 8-59
8-15-3 Preparations
1.) Power Off/Shutdown the system.
!! CAUTION:
Before you replace the boards, please make sure you wear the Ground strap, which prevents
ESD problems just in case.