GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-25
19.)Go to “Save & Exit” in BIOS menu and select “Save Changes and Exit”. Then, the system will re-
boot
8-3-7-3 Application Software Installation Preparation
1.) Verify the system boot up to Windows Screen
2.) Prepare the Application Software DVD.
3.) Press DVD-Drive eject button, and place the Application Software DVD on the tray.
4.) Click Windows Explorer icon on the taskbar.