EasyManuals Logo

GE LOGIQ S7 Pro User Manual

GE LOGIQ S7 Pro
435 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #57 background imageLoading...
Page #57 background image
GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 2 - Site Preparation 2-5
2-2-4 Probe Environmental Requirements
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Table 2-5 2D Probe Environmental Requirements
Operational
Temperature
10
o
- 35
o
C
for 2D probe
18
o
- 30
o
F
for 4D probe
Humidity
10 - 80% (non-condensing) for 2D probe
30 - 80% (non-condensing) for 4D probe
Pressure 700 - 1060hPa
NOTICE
!! NOTICE:
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
0 TO + 55 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
CAUTION
!! CAUTION:
Have two people available to deliver and unpack the LOGIQ™ S7 ultrasound system.
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
result in personal injury and/or damage to the system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S7 Pro and is the answer not in the manual?

GE LOGIQ S7 Pro Specifications

General IconGeneral
BrandGE
ModelLOGIQ S7 Pro
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals