GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-8 Section 8-3 - System Software - Installation/Upgrade Procedure
- General, Record preset Region, Language and Regional Options. (See the latest revision of
the LOGIQ S7 Expert/Pro Basic User Manual)
8-3-5-2-1-2 Connectivity - Recording the TCP/IP settings
1.) From the touch panel, select Utility -> Connectivity -> TCP/IP.
Figure 8-1 Connectivity - Recording the TCP/IP settings
2.) Record all settings in Table 8-3.
8-3-5-2-1-3 Connectivity — WLAN
1.) Press Utility--> Connectivity--> TCP/IP--> Wireless Network--> Configuration. The Wireless
Network Configuration tool appears. Available Wireless Networks appear.
2.) Select the Properties Tab.
3.) Record all settings in Table 8-4.
Table 8-3 Record settings from TCP/IP screen
Parameter Value
Computer Name
IP Settings
Enable DHCP ON / OFF
IP-Address
Subnet Mask
Default Gateway
Network Speed