GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-22 Section 8-3 - System Software - Installation/Upgrade Procedure
boot.
6.) Press DVD-Drive eject button.
7.) Place the Base System Software DVD on the tray.
8.) Wait few moments for DVD drive to read DVD contents.
9.) The notification screen appears, and then hit the Enter key.
10.)Select either
[A] “Load the complete disk”, or
[B] "Load the bootable C: and repository Z: partition only"
and hit the Enter key.