GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-95
3.) Unscrew five screws to separate SOM board from the GFS T6.
4.) Pull the SOM board out of the GFS T6..
Figure 8-104 Unscrewing to separate SOM board
Figure 8-105 Drawing up the SOM board
!! CAUTION:
When replacing with new SOM, Push the SOM in both side of GFS T6 Assy like as below Figure 8-106.
Figure 8-106 Push the SOM in both side of GFS Assy