GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-132 Section 8-31 - Replacement of the Monitor Arm
5.) Unscrew 1 screw and remove the 1st arm cover polaris. Refer to the figure below.
6.) Remove 1 cable ties and extract the monitor cable through the hole. Refer to the figure below.
7.) Unscrew one screw to extract the setscrew. Refer to the figure below.
Figure 8-154 Removing the 1st arm cover
Figure 8-155 Remove cable ties and pull out Harness cable
Figure 8-156 Extracting the setscrew