GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-33
Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Type BF Patient Leakage Current Test - Mains to Probe (This test should be done for all probes)
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
Table 10-23
Tester
Neutral
Tester Polarity
Switch
Ground
Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 0.1mA
Closed Normal Open 0.5mA
Closed Reversed Closed 0.1mA
Closed Reversed Open 0.5mA
Open Normal Closed 0.5mA
Open Reversed Closed 0.5mA
Table 10-24
Tester
Neutral
Tester Polarity
Switch
Ground
Switch
Max Value
Allowed
Value
Measured
OK? Comments
Closed Normal Closed 5mA
Closed Reversed Closed 5mA