8. 14. 6. 2 Stitching
Stitching Adjustment is to calibrate the amount of shift of each image block scanned by CIS in
order to organize 5 pieces of image blocks into 1 complete image, based on a designated
calibration chart “Shading Sheet”.
One sheet of “Shading Sheet” is included in every scanner accessory.
1. Run K129 Diag. Click [BackupData] to recall “Backup Data” list sub window.
NOTE
BUD No.15 (stitch setting 1) should be temporarily set to OFF “0” during Stitching
Adjustment.