Remote Control Commands
R&S
®
SMW200A
1030User Manual 1175.6632.02 ─ 30
Parameters:
<ArbSignalType> SINE | RECT | CIQ | AWGN
*RST: SINE
Example: See Example "Creating test signals" on page 1019.
Options: AWGN requires R&S SMW-K811
Manual operation: See "Test Signal Form" on page 306
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:CIQ:I <I>
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:CIQ:Q <TSig>
Sets the value for the I and Q component of the test signal
Parameters:
<TSig> float
Range: -1 to 1
Increment: 0.001
*RST: 0
Default unit: FS
Example: See Example "Creating test signals" on page 1019
Manual operation: See "I Value, Q Value" on page 310
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:AMPLitude <Amplitude>
Sets the digital amplitude of the rectangular wave.
Parameters:
<Amplitude> float
Range: 0 to 1
Increment: 0.001
*RST: 0.800
Default unit: FS
Example: see Example "Creating test signals" on page 1019
Manual operation: See "Amplitude" on page 309
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:FREQuency <Frequency>
Sets the frequency of the test signal.
Parameters:
<Frequency> float
Range: 100 to depends on the installed options
Increment: 0.01
*RST: 1000
Default unit: Hz
Example: See Example "Creating test signals" on page 1019.
SOURce Subsystem