[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal]:DELay............................................... 1065
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal]:TDELay............................................. 1066
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal]:RDELay?...........................................1066
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal]:INHibit............................................... 1067
[:SOURce<hw>]:BB:ARBitrary[:TRIGger]:SEQuence <Sequence>
The command selects the trigger mode.
Parameters:
<Sequence> AUTO | RETRigger | AAUTo | ARETrigger | SINGle
*RST: AUTO
Example: See Example "Managing waveform files" on page 1020
Manual operation: See " Mode" on page 253
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SOURce
<Source>
Selects the trigger signal source and determines the way the triggering is executed.
Provided are:
●
Internal triggering by a command (INTernal)
●
External trigger signal via one of the local or global connectors
– EGT1|EGT2: External global trigger
– EGC1|EGC2: External global clock
– ELTRigger: External local trigger
– ELCLock: External local clock
●
Internal triggering by a signal from the other basebands (INTA|INTB)
●
In primary-secondary instrument mode, the external baseband synchronization sig-
nal (BBSY)
●
OBASeband|BEXTernal|EXTernal: Setting only
Provided only for backward compatibility with other Rohde & Schwarz signal gener-
ators.
The R&S SMW accepts these values and maps them automatically as follows:
EXTernal = EGT1, BEXTernal = EGT2, OBASeband = INTA or INTB
(depending on the current baseband)
Parameters:
<Source> INTB|INTernal|OBASeband|EGT1|EGT2|EGC1|EGC2|ELTRig-
ger|INTA|ELCLock|BEXTernal|EXTernal | BBSY
*RST: INTernal
Example: See Example "Managing waveform files" on page 1020
Options: ELTRigger|ELCLock require R&S SMW-B10
BBSY require R&S SMW-B9
Manual operation: See "Source" on page 254