Loading...
Remote Control Commands
R&S
®
SMW200A
932User Manual 1175.6632.02 ─ 30
Example: See Example "Connecting and configuring external instrument"
on page 905.
Manual operation: See "External Instrument" on page 145
See "Apply" on page 148
See "Apply and Connect" on page 149
:SCONfiguration:EXTernal:CODer<ch>:REMote:CONNect
:SCONfiguration:EXTernal:FADer<ch>:REMote:CONNect
:SCONfiguration:EXTernal:BBMM<ch>:REMote:CONNect
:SCONfiguration:EXTernal:RF<ch>:REMote:CONNect
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:CONNect
Triggers the connection establishment.
Example:
See Example "Connecting and configuring external instrument"
on page 905.
Usage: Event
Manual operation: See "Apply and Connect" on page 149
:SCONfiguration:EXTernal:CODer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:FADer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:BBMM<ch>:REMote:INFO?
:SCONfiguration:EXTernal:RF<ch>:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INFO?
Queries information on the external instrument.
Return values:
<Info> string
Returns information in one of the following formats:
"None"
An external instrument is not assigned
<SymbolicName>,<RemoteChannel>,<RemoteAd-
dress>,<RF Path>
An external instrument is assigned, but not connected
<SymbolicName>,<RemoteChannel>,<RemoteAd-
dress>,<RF Path>[,<*IDN? String>]
An external instrument is assigned and connected
Example:
:SCONfiguration:EXTernal:BBMM1:REMote:INFO?
"SGT (101676)","LAN","rssgt100a101676","A",
"Rohde&Schwarz,SGT100A,1419.4501k02/101676,3.1.19.4-3.18.251.99"
Usage: Query only
SCONfiguration Subsystem