Remote Control Commands
R&S
®
SMW200A
950User Manual 1175.6632.02 ─ 30
Example: See Example "Global connectors settings configuration"
on page 945.
Options: SYNCIN requires R&S SMW-B9
BERDATIN|BERCLKIN|BERDATENIN|BERRESTIN require
R&S SMW-K80
ERRTA|ERRTB require R&S SMW-K78
Manual operation: See "Signal" on page 725
:OUTPut:USER<ch>:SIGNal
<Signal>
Sets the control signal that is output at the selected connector.
To define the connector direction, use the command :OUTPut:USER<ch>:
DIRection.
Suffix:
USER<ch>
.
1 to 6
Parameters:
<Signal> MARKA1 | MARKA2 | MARKA3 | MARKB1 | MARKB2 |
MARKB3 | MARKC1 | MARKC2 | MARKC3 | MARKD1 |
MARKD2 | MARKD3 | SVALA | SVALB | OPULSA | OPULSB |
SYNCA | VIDEOA | VIDEOB | SYNCB | NONE | RTRIGA |
RTRIGB | SVALANegated | SVALBNegated | LOW | HIGH |
MTRigger | SYNCOUT | BERRESTOUT | BERDATENOUT |
BERCLKOUT | BERDATOUT
MARK<A|B|C|D><1|2|3> = Baseband <BB> Marker 1/2/3
(available marker signals depend on the system configuration,
see Table 12-1)
SVALA|SVALB = Signal Valid A/B, available for USER4|5|6
SVALANegated|SVALBNegated = Signal Valid A/B (negative),
available for USER4|5|6
OPULSA|OPULSB = Pulse Out A/B, available for USER4|5|6
SYNCA|SYNCB = Pulse Sync A/B, available for USER4|5|6
VIDEOA|VIDEOB = Pulse Video A/B, available for USER4|5|6
MTRigger = Manual Trigger, available for USER6
RTRIGA|RTRIGB = REG trigger A/B, available for USER4|5
BERRESTOUT|BERDATENOUT|BERCLKOUT|BERDATOUT =
BERT TestGen Data, Clock, Data Enable and Restart
SYNCOUT = Baseband Sync Out
LOW|HIGH = Always 0/1
NONE = none
*RST:
MARKA1
Example: See Example "Global connectors settings configuration"
on page 945
Options: SYNCOUT requires R&S SMW-B9
SOURce Subsystem