Applications
R&S
®
RTM3000
172User Manual 1335.9090.02 ─ 09
Width X
Changes the width of the mask in horizontal direction. The specified factor in divisions
is added to the positive x-values and subtracted from the negative x-values of the
mask limits in relation to the mask center. Thus, the left half of the mask is pulled to the
left, the right half is pulled to the right.
Remote command:
MASK:XWIDth on page 537
Width Y
Changes the width of the mask in vertical direction. The specified number of divisions
is added to the y-values of the upper mask limit and subtracted from the y-values of the
lower mask limit. Thus, the upper half of the mask is pulled upwards, the lower half is
pulled down, and the overall height of the mask is twice the "Width Y".
Remote command:
MASK:YWIDth on page 537
Save
Saves the mask in an instrument-specific format (MSK). The complete mask definition
- envelope waveform with width, stretch and position settings - is stored.
Remote command:
MASK:SAVE on page 537 and commands described in Chapter 17.9.1.5, "Masks",
on page 577.
Load Mask
Opens a file explorer to select a previously stored mask. The selected mask is loaded
and can be used for a subsequent test.
Remote command:
MASK:LOAD on page 538
Actions
Opens a submenu to select the actions to be taken when a violation against the mask
limits occurs.
"Sound"
Generates a beep sound at each violation.
"Stop"
Stops the waveform acquisition. Set the number of the stop violation.
Thus, you can ignore a number of violations before stop.
"Pulse"
Creates a pulse on the [Aux Out] connector. This selection sets the
configuration of the [Aux Out] connector to "Mask Violation".
"Screenshot"
Saves a screenshot according to the settings in "File" > "Screen-
shots".
"Save Wave-
form"
Saves the waveform data according to the settings in "File" > "Wave-
forms".
Remote command:
MASK:ACTion:SOUNd:EVENt:MODE on page 538
MASK:ACTion:STOP:EVENt:MODE on page 539
MASK:ACTion:STOP:EVENt:COUNt on page 539
MASK:ACTion:PULSe:EVENt:MODE on page 538
Mask testing

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals