Remote control commands
R&S
®
RTP
1200User Manual 1337.9952.02 ─ 12
DC
Direct connection with 50 Ω termination, passes both DC and
AC components.
DCLimit
Direct connection with 1 MΩ termination, passes both DC and
AC components.
Only available if the 1 M adapter is used.
AC
Connection with 1 MΩ termination, removes DC and very low-
frequency components.
Only available if the 1 M adapter is used.
*RST:
DC
CHANnel<m>:SCALe <Scale>
Sets the vertical scale for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Scale> Scale value, given in Volts per division.
Range: Depends on attenuation factors, coupling, and
instrument model, see below.
Increment: Depends on vertical and probe settings
*RST: 0.05
Default unit: V/div
With 1:1 probe and external attenuation = 1, the vertical scale
(input sensitivity) is minimum 2 mV/div (1 mV/div in high defini-
tion mode). If the probe and/or external attenuation is changed,
multiply the values by the attenuation factors to get the actual
scale range.
Usage:
Asynchronous command
CHANnel<m>:RANGe <Range>
Sets the voltage range across the 10 vertical divisions of the diagram. Use the com-
mand alternativly instead of CHANnel<m>:SCALe.
Suffix:
<m>
.
1..4
Selects the input channel.
Acquisition and setup