Remote control commands
R&S
®
RTP
1223User Manual 1337.9952.02 ─ 12
D16 = 1.6 mm
D25 = 2.5 mm
D45 = 4.5 mm
D75 = 7.5 mm
For R&S RT-ZMA40:
NONE = no submodule
SRCS = semi-rigid cables
Usage: Asynchronous command
PROBe<m>:DEEMbedding:BANDwidth <ProbeBandwidth>
Sets the maximum bandwidth until which the signal is corrected. This maximum value
is the minimum bandwidth value of probe, probe tip and oscilloscope bandwidth, and
can not be higher than the highest frequency in a used S-parameter file. Consider that
most tip modules support the full bandwidth of the probe amplifier, but some tip mod-
ules have limited bandwidth.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<ProbeBandwidth> Range: 1E+9 to 16E+9
Increment: 10
*RST: 4E+9
Default unit: Hz
Usage: Asynchronous command
PROBe<m>:DEEMbedding:MODE <Mode>
Sets the probing mode.
Suffix:
<m>
.
1..4
Selects the channel to which the probe is connected.
Parameters:
<Mode> PROBing | TERMinating
PROBing
For measurements with a high-ohmic probe and measure on an
existing line in parallel to the load.
TERMinating
For measurements if the measuring equipment is the load of the
line.
*RST: PROBing
Usage: Asynchronous command
Acquisition and setup