Remote control commands
R&S
®
RTP
1561User Manual 1337.9952.02 ─ 12
TRIGger<m>:I2C:DPOPerator <DataPosOperator>
Sets the operator for the data position. You can defined an exact position, or a position
range.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
ANY = OFF
The position of the required pattern is not relevant.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with TRIGger<m>:I2C:DPOSition.
INRange = RANGe
In range: Set the minimum and maximum value of the range with
TRIGger<m>:I2C:DPOSition and TRIGger<m>:I2C:DPTO.
*RST: ANY
TRIGger<m>:I2C:DPOSition
<DataPosition>
Sets the number of data bytes before the first byte of interest. These bytes are ignored.
Parameters:
<DataPosition> The index 0 is associated with the first data byte.
Range: 0 to 4095
Increment: 1
*RST: 0
TRIGger<m>:I2C:DPTO <DataPositionTo>
Defines the last byte of interest, if TRIGger<m>:I2C:DPOPerator is set to RANGe.
Parameters:
<DataPositionTo> Range: 0 to 4095
Increment: 1
*RST: 0
TRIGger<m>:I2C:DCONdition <DataOperator>
Sets the operator to set a specific data value or a data range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
*RST: EQUal