Remote control commands
R&S
®
RTP
1574User Manual 1337.9952.02 ─ 12
SEARch:TRIGger:I2C:AMODe <SearchName>,<AddressType>
SEARch:TRIGger:I2C:AMODe? <SearchName>
Sets the address length.
Parameters:
<AddressType> BIT7 | BIT7_RW | BIT10 | ANY
BIT7 | BIT10
Enter only address bits in the address pattern.
BIT7_RW
Enter seven address bits and also the R/W bit in the address
pattern.
ANY
Only available for search criteria "Address and data" (SEARch:
TRIGger:I2C:ADData is set ON). Used to search for data only,
regardless of the address.
*RST:  BIT7
Parameters for setting and query: 
<SearchName>
Usage:  Asynchronous command
SEARch:TRIGger:I2C:ADDRess <SearchName>,<Address>
SEARch:TRIGger:I2C:ADDRess? <SearchName>
Specifies an address pattern, or sets the the start value of an address range.
Parameters:
<Address> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148.
The pattern length is defined with SEARch:TRIGger:I2C:
AMODe.
Parameters for setting and query: 
<SearchName>
SEARch:TRIGger:I2C:ADDTo <SearchName>,<AddressTo>
SEARch:TRIGger:I2C:ADDTo? <SearchName>
Sets the the end value of an address range if SEARch:TRIGger:I2C:ACONdition
is set to INRange or OORange.
Parameters:
<AddressTo> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148.
The pattern length is defined with SEARch:TRIGger:I2C:
AMODe.
Parameters for setting and query: 
<SearchName>
Protocols