EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1618

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1618User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Bitrate> Range: 100 to 5E+6
Increment: 1
*RST: 100E+3
Default unit: bps
BUS<m>:CAN:FDATa:ENABle <CANFDEnabled>
Enables the CAN FD protocol configuration.
The setting is available in CAN FD option R&S RTP-K9.
Suffix:
<m>
.
1..4
Parameters:
<CANFDEnabled> ON | OFF
*RST: ON
BUS<m>:CAN:FDATa:DBITrate <FlexDatBitrate>
Sets the bit rate of the data phase. The data rate can be higher than the arbitration
rate, but it is uniform and fixed for a given CAN FD bus.
The setting is available in CAN FD option R&S RTP-K9.
Suffix:
<m>
.
1..4
Parameters:
<FlexDatBitrate> Range: 100 to 15E+6
Increment: 1
*RST: 1E+6
Default unit: bps
BUS<m>:CAN:FDATa:SAMPlepoint <FlexDatSmpPt>
BUS<m>:CAN:SAMPlepoint <SamplePoint>
Sets the position of the sample point within the bit in percent of the nominal bit time.
Alternatively, you can set the sample point with BUS<m>:CAN:T1Segment and
BUS<m>:CAN:T2Segment.
For CAN FD signals, BUS<m>:CAN:SAMPlepoint defines the synchronization of the
arbitration phase, and BUS<m>:CAN:FDATa:SAMPlepoint defines the synchroniza-
tion of the data phase.
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals