Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<FlexDatSmpPt>
<SamplePoint>
Range: 12 to 96 if number of segments is 24. The range
reduces if the signal has fewer segments.
Increment: 5
*RST: 66
Default unit: %
BUS<m>:CAN:FDATa:T1Segment
<FlexDatTimeSeg1>
BUS<m>:CAN:T1Segment <TimeSeg1>
Sets the number of time quanta before the sample point (T1Segment). T1Segment
comprises the segments Synch_seg, Prop_seg, and Phase_seg1 which are specified
in the CAN standard.
Make sure to set also BUS<m>:CAN:T2Segment for correct definition of the sample
point. Alternatively, you can use BUS<m>:CAN:SAMPlepoint.
See also: Chapter 13.5.2.2, "Advanced settings", on page 559
For CAN FD signals, BUS<m>:CAN:T1Segment defines the synchronization of the
arbitration phase, and BUS<m>:CAN:FDATa:T1Segment defines the synchronization
of the data phase.
Suffix:
<m>
.
1..4
Parameters:
<FlexDatTimeSeg1>
<TimeSeg1>
Time quanta
Range: 3 to 23
Increment: 1
*RST: 6.6
BUS<m>:CAN:FDATa:T2Segment <FlexDatTimeSeg2>
BUS<m>:CAN:T2Segment <TimeSeg2>
Sets the number of time quanta after the sample point (T2Segment). T2Segment
matches Phase_seg2 specified in the CAN standard.
Make sure to set also BUS<m>:CAN:T1Segment on page 1619 for correct definition of
the sample point. Alternatively, you can use BUS<m>:CAN:SAMPlepoint.
See also: Chapter 13.5.2.2, "Advanced settings", on page 559
For CAN FD signals, BUS<m>:CAN:T2Segment defines the synchronization of the
arbitration phase, and BUS<m>:CAN:FDATa:T2Segment defines the synchronization
of the data phase.