Remote control commands
R&S
®
RTP
1660User Manual 1337.9952.02 ─ 12
SYNC
Start of the frame, triggers on the stop bit of the sync field.
ID
Sets the trigger to one specific identifier or an identifier range.
To set the identifier, use TRIGger<m>:LIN:ICONdition,
TRIGger<m>:LIN:IMIN on page 1661, and TRIGger<m>:
LIN:IMAX on page 1661.
IDOR
Triggers on an OR combination with up to four identifier condi-
tions. For each identifier condition, enable it with TRIGger<m>:
LIN:IDOR<n>:ENABle and set the value with TRIGger<m>:
LIN:IDOR<n>[:VALue]
IDDT
Combination of identifier and data conditions
To set the identifier condition, use TRIGger<m>:LIN:
ICONdition, TRIGger<m>:LIN:IMIN, and TRIGger<m>:
LIN:IMAX.
To set the data condition, use TRIGger<m>:LIN:DCONdition,
TRIGger<m>:LIN:DMIN, and TRIGger<m>:LIN:DMAX.
WKFR
Wakeup frame
ERRC
Error condition. Define the error types with TRIGger<m>:LIN:
CHKSerror on page 1664, TRIGger<m>:LIN:IPERror, and
TRIGger<m>:LIN:SYERror
*RST:
SYNC
TRIGger<m>:LIN:ICONdition <IdOperator>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These condition require one identifier pat-
tern to be set with TRIGger<m>:LIN:IMIN
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:LIN:IMIN and TRIGger<m>:
LIN:IMAX
*RST: EQUal
Protocols