Remote control commands
R&S
®
RTP
1732User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
Parameters:
<CRCOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater Than,
Greater than or equal. These conditions require a destination
address to be set with TRIGger<m>:ETHernet:FRAMe:CMIN.
INRange | OORange
In range, Out of range. Set the minimum and maximum value of
the range with TRIGger<m>:ETHernet:FRAMe:CMIN and
TRIGger<m>:ETHernet:FRAMe:CMAX.
*RST: EQUal
TRIGger<m>:ETHernet:FRAMe:CMAX
<CRCPatternMax>
Sets the end value of the frame check sequence range, if TRIGger<m>:ETHernet:
FRAMe:CCONdition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<CRCPatternMax>
TRIGger<m>:ETHernet:FRAMe:CMIN <CRCPatternMin>
Specifies a CRC pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<CRCPatternMin>
TRIGger<m>:ETHernet:FRAMe:DCONdition <DestAddrOptor>
Sets the destination address operator to trigger on a specific pattern or a range.
Suffix:
<m>
.
1..3
Parameters:
<DestAddrOptor> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater Than,
Greater than or equal. These conditions require a destination
address to be set with TRIGger<m>:ETHernet:FRAMe:DMIN.