Remote control commands
R&S
®
RTP
1796User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
Parameters:
<TypeOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:TBTO:LENGth:
MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:TBTO:LENGth:MIN and
TRIGger<m>:TBTO:LENGth:MAX.
*RST:
EQUal
TRIGger<m>:TBTO:LENGth:MAX
<TypePatternMax>
Sets the end value of the length / type range if TRIGger<m>:TBTO:LENGth:
CONDition is set to INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<TypePatternMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:TBTO:LENGth:MIN <TypePatternMin>
Specifies the length / type bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<TypePatternMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:TBTO:SADDress:CONDition <SrcAddrOperator>
Sets the condition for the source address. You can define an exact value or a value
range.