Remote control commands
R&S
®
RTP
1999User Manual 1337.9952.02 ─ 12
SEARch:TRIGger:MDIO:FRAMetype <SearchName>,<FrameType>
SEARch:TRIGger:MDIO:FRAMetype? <SearchName>
Allows to select the Type of Frame code (or OP code, OpCode, operation code); avail-
able only in search criteria type "Data".
Note that the same OpCode may have different meanings in Clause 22 and Clause 45.
Parameters:
<FrameType> OP00 | OP01 | OP10 | OP11 | OPXX
OP00
Address frame (in Clause 45, only)
OP01
Write frame (in Clause 22 or Clause 45)
OP10
Read frame (in Clause 22) or Post-Read increment address
frame (in Clause 45)
OP11
Read frame (in Clause 45)
OPXX
Any frame type
*RST:
OPXX
Note that the user interface shows interpretations of the numeri-
cal OpCode values corresponding to Clause 45. Clause 22 is
not represented by this interpretation.
Also, note that OPXX will never be a result of decoding, but it is
still an option for triggering.
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MDIO:PHYS <SearchName>,<PhyAddr>
SEARch:TRIGger:MDIO:PHYS? <SearchName>
Allows to set the physical address (in Clause 22) or port address (in Clause 45) of the
frame pattern (5 bits); available only in search criteria type "Data".
Parameters:
<PhyAddr>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MDIO:REGI <SearchName>,<RegAddr>
SEARch:TRIGger:MDIO:REGI? <SearchName>
Allows to set the register address (in Clause 22) or device address (in Clause 45) of
the frame pattern (5 bits); available only in search criteria type "Data".