Remote control commands
R&S
®
RTP
2064User Manual 1337.9952.02 ─ 12
TRIGger<m>:USBThree:DOPerator <Frame>, <Field>, <Operator>
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:DOPerator <DataOperator>
Sets the operator for the data pattern in the selected field of the selected frame.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n> *
Specifies the field number within the frame.
<o> *
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one endpoint
value to be set using TRIGger<m>:USBThree:FRAMe<n>:
FLD<o>:DMIN.
INRange | OORange
In range, out of range. These conditions require a range of end-
point values to be set using TRIGger<m>:USBThree:
FRAMe<n>:FLD<o>:DMIN and TRIGger<m>:USBThree:
FRAMe<n>:FLD<o>:DMAX.
*RST:
EQUal
TRIGger<m>:USBThree:DMIN <Frame>, <Field>, <Data>
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:DMIN <DataMin>
Specifies the data pattern, or sets the start value of a data pattern range.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n> *
Specifies the field number within the frame.
<o> *
Parameters:
<DataMin>
TRIGger<m>:USBThree:DMAX <Frame>, <Field>, <Data>
TRIGger<m>:USBThree:FRAMe<n>:FLD<o>:DMAX <DataMax>
Sets the end value of a data pattern range if TRIGger<m>:USBThree:FRAMe<n>:
FLD<o>:DOPerator is set to INRange or OORange.