Remote control commands
R&S
®
RTP
2247User Manual 1337.9952.02 ─ 12
Parameters:
<OutputDomain> TIME | MDIStance | INDistance
MDIStance = distance in m
IDIStance = distance in inch
*RST:  TIME
Usage:  Asynchronous command
TDRT:PERMittivity <Permittivity>
Sets the permittivity "ε_r" of the used DUT.
Parameters:
<Permittivity> Range:  1  to  100
Increment:  0.1
*RST:  3.3
Usage:  Asynchronous command
TDRT:SIGNal<m>STATe <State>
Enables the indicated TDR/TDT signal.
Suffix: 
<m>
.
1..3
1 = impedance Z (result of TDR measurement)
2 = reflection coefficient ρ (result of TDR measurement)
3 = step response h (result of TDT measurement)
Parameters:
<State> ON | OFF
*RST:  OFF
Usage:  Asynchronous command
TDRT:SIGNal<m>:SCALe <VerticalScale>
Set the vertical scale of the indicated TDR/TDT signal.
Suffix: 
<m>
.
1..3
See TDRT:SIGNal<m>STATe on page 2247.
Parameters:
<VerticalScale> Range:  1E-12  to  100E+12
Increment:  1E-3
*RST:  0.05
Default unit: /div
Usage:  Asynchronous command
Pulse source and TDR/TDT analysis (options R&S RTP-B7/K130)